The is a high-performance, dual-wavelength UV/Vis detector designed for use with Shimadzu’s Prominence and LC-20A series liquid chromatography systems. This user manual serves as the definitive technical guide for laboratory personnel responsible for installation, operation, maintenance, and troubleshooting of the detector.
Plug in the unit, turn it on, and reset the lamp usage hour counter via the utility menu on the control panel. 6. Troubleshooting Guide Probable Cause Recommended Action Blown fuse or disconnected power cable.
Refer to the for detailed procedures.
Comprehensive Guide to the Shimadzu SPD-20A Go to product viewer dialog for this item. User Manual: Operation, Maintenance, and Troubleshooting The Shimadzu SPD-20A Go to product viewer dialog for this item. shimadzu spd20a user manual
The service manual outlines procedures for calibrating the optical system, including adjusting filters and mirrors to ensure accurate wavelength detection. 5. Safety and Troubleshooting The emphasizes safety:
These are the top questions that the answers directly.
The Shimadzu SPD-20A is a versatile UV-Vis detector designed for use in high-performance liquid chromatography (HPLC) systems. Its advanced design and cutting-edge technology enable it to deliver precise and reliable absorbance measurements in the ultraviolet and visible spectrum ranges. The instrument features a large, easy-to-read LCD display, intuitive navigation, and a robust design that withstands the demands of daily laboratory use. Comprehensive Guide to the Shimadzu SPD-20A Go to
: The "AV" version allows both deuterium (D2) and tungsten (W) lamps to be lit at once, enabling high-sensitivity detection across both UV and visible light ranges Documentation & Maintenance Resources
An error has occurred (often a lamp failure or communication break).
190 nm to 700 nm (190 nm to 900 nm for the SPD-20AV). Spectral Bandwidth: 8 nm. Wavelength Accuracy: ±1 nm. Wavelength Reproducibility: ±0.1 nm. Noise Level: AU (under specified standard conditions). Drift: AU/hour (under specified standard conditions). 2. Installation and Initial Setup The system logs cumulative usage
Ensure the unit is properly grounded. According to the instruction manual, connect the power cable securely and turn the power ON.
Avoid placing the instrument directly under air conditioning vents or in direct sunlight, as temperature fluctuations cause baseline drift. Fluidic and Electrical Connections
Finally, lamp management is crucial. The D2 lamp has a finite lifespan (typically 1,000 to 2,000 hours). The system logs cumulative usage, and operators are advised to keep spare lamps on hand to avoid downtime. When replacing the lamp, the manual strictly warns against touching the bulb glass with bare fingers, as oils from the skin can create hot spots and cause premature failure.
Remove the front covers/screws securing the lamp housing access panel. Disconnect the lamp power connector.
Connect the waste tubing to the flow cell outlet port (typically the top port).